Description
Product Details
The YESAIR uses a remote sample probe that enters through the port on the door or side of the incubator and provides a direct reading of the CO2 (and O2 if installed) levels inside the chamber. Coupled with our 3-point calibration standard, and the use of an infrared sensor, the CO2 sensor offers stable and accurate measurement readings of the CO2 levels.
Incubators are maintained at fairly high temperatures and are required to sustain a high relative humidity. The YESAIR has built in sensors at the tip of the secondary probe to sample temperature and RH to confirm the internal chamber readings.
Includes monitor with pump, remote sampling probe to measure target gas, remote sampling probe to measure temperature and RH, calibration tubing, three “AA” NiMH rechargeable batteries, AC adapter and battery charger in a hard plastic carrying case. The optional YES Logger package is available with 2GB memory card, card reader, USB cable and YES Viewer Software.
Features
- Stable infrared CO2 sensor technology, available with O2 sensor also
- Remote sample probes for easy entry into incubator port
- Large graphic LCD display with a multi-function, easy to use menu
- Light weight, contoured and comfortable to hold
- RoHS compliant circuit boards
Applications
- Hospitals
- Laboratories
- Poultry incubators
- Infant incubators
- IVF Facilities
- Bacteriological incubators
- Biomedical research facilities